Remote control commands
R&S
®
RTE
1703User Manual 1326.1032.02 ─ 20
Parameters:
<DataSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST: C1W1
Usage: Asynchronous command
BUS<m>:SENT:DATA:THReshold <Threshold>
Sets a user-defined threshold value. Alternatively, you can set the threshold according
to the signal technology BUS<m>:SENT:TECHnology.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<Threshold> Range: -12 to 12
Increment: 0.1
*RST: 1.65
Default unit: V
BUS<m>:SENT:TECHnology <Technology>
Selects the threshold voltage.
Suffix:
<m>
.
1..4
Selects the serial data bus.
Parameters:
<Technology> V25 | MANual
V25
The threshold value is 2.5 V, according to CMOS technology.
MANual
Sets the threshold to the value set with BUS<m>:SENT:DATA:
THReshold.
*RST: V25
BUS<m>:SENT:CLKPeriod <ClockPeriod>
Sets the nominal clock period (clock tick).