Remote control commands
R&S
®
RTE
1704User Manual 1326.1032.02 ─ 20
Parameters:
<ClockPeriod> Range: 1E-6 to 100E-6
Increment: 1E-6
*RST: 6E-6
Default unit: s
BUS<m>:SENT:CLKTolerance <ClockTolerance>
Sets a tolerated deviation of the clock signal.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ClockTolerance> Range: 0 to 25
Increment: 1
*RST: 20
Default unit: %
BUS<m>:SENT:DNIBbles <DataNibbles>
Sets the number of data nibbles for a transmission sequence.
Suffix:
<m>
.
1..4
Selects the serial data bus.
Parameters:
<DataNibbles> Range: 1 to 6
Increment: 1
*RST: 6
BUS<m>:SENT:SFORmat <Format>
Selects the serial message format.
Suffix:
<m>
.
1..4
Parameters:
<Format> SHORt | ENHanced | NONE
Short serial message, Enhanced serial message, none = single
transmission sequence.
*RST: NONE
BUS<m>:SENT:CRCVersion <CRCVersion>
Selects the calculaction method for the cyclic redundancy check (CRC).
Protocols