Protocol analysis
R&S
®
RTE
740User Manual 1326.1032.02 ─ 20
"Serial Message"
Triggering on a serial message depends on the serial protocol
selected with Serial protocol > Short | Enhanced and the associated
setting parameters:
●
"Short" serial message
– "Sequence > Identifier": triggers at the end of the "ID".
– "Sequence > ID+Data": triggers at the end of the "ID and
Data".
Description of trigger type specific settings: "Serial Message
setup" on page 742
●
"Enhanced" serial message
– "Sequence > Identifier": triggers at the end of the "ID".
– "Sequence > ID+Data": triggers at the end of the "ID and
Data".
Description of trigger type specific settings: "Serial Message
setup" on page 742
"Error condition"
Triggers if certain errors occur.
You can select the following error events for triggering:
●
"Form Error"
●
"Calibration Pulse Error"
●
"Pulse Period Error"
●
"CRC Error"
●
"Irregular Frame Length Error"
Description of error-specific trigger conditions, see "Error conditions
setup" on page 744.
Remote command:
TRIGger<m>:SENT:TYPE on page 1706
Transmission Sequence setup
Configures the trigger conditions for trigger type transmission sequence.
SENT (option R&S
RTE-K10)