Instrument Function
R&S
®
SMU200A
270Operating Manual 1007.9845.32 ─ 15
In list operating modes Step and Extern Step, the set dwell time does not affect signal
generation. In this case, the duration of a list step is determined by the time between two
(internal or external) trigger events.
The "Dwell Time"
set by the user is used as the step time of the list mode. The effective
net dwell time is shorter, reduced by the setting time. This setting time may be greater
than the time specified in the data sheet.
SCPI command:
[:SOURce<hw>]:LIST:DWELl on page 753
Current Index - List Mode
Sets the list index in "Step" mode.
SCPI command:
[:SOURce<hw>]:LIST:INDex on page 754
Blank RF Output (Default ON) - List Mode
Activates/deactivates RF output blanking. If activated, blanking occurs on all frequency
changes. If deactivates, blanking is always disabled.
SCPI command:
:OUTPut<hw>:BLANk:LIST:STATe on page 554
Learn List Mode Data... - List Mode
Starts the determination of the hardware setting for the selected list. The data determined
in this way is stored along with the list.
It may be necessary to deliberately activate list learning in the event of greatly altered
environmental conditions that require new hardware settings.
If this is not done, a previously learned hardware setting will continue to be used when
list mode is switched on ("State = On"). If no setting is available, e.g. when the list is used
for the first time, learning is automatically activated.
SCPI command:
[:SOURce<hw>]:LIST:LEARn on page 756
List Mode Data... - List Mode
Calls the "File Select" menu for selecting and creating a list or the "File Manager".
RF Signal and Analog Modulations - RF Block