Examination
ARCADIS Varic
SPR2-310.620.30.02.02 35
of 52
Native images from the database 0
Native images from previous examinations acquired in Subtraction mode (op-
tion) or Roadmap mode (option) are stored in the local database or in your archive
together with the subtraction images. Both image types can be displayed simul-
taneously on both monitors.
Loading and
displaying images
â—† Load the images of the requested patient into the Viewing task card by using
the Patient Browser.
â—† Select Sub > Sub/Native Display in the main menu of the Viewing task card.
—or—
â—† Click this button in the Sub subtask card.
– In the Viewing task card the subtraction images are displayed on the left
monitor.
– In the Native task card the corresponding native images are displayed on
the right monitor.
Closing the patient After finishing your evaluation of patient images, close the folder of this patient.
Now you can load images of the next patient.
â—† Select Patient > Close Patient in the main menu of the Viewing task card.
– The images in the Viewing and Native task cards are unloaded. Both task
cards are empty now.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors