Maintenance
Checking the EMERGENCY STOP function for motor-driven
system movements
0
â—† Switch the ARCADIS Varic on.
â—† Move the lifting column and press the EMERGENCY STOP button at the
same time.
– The lifting movement is stopped.
– A message is displayed on the monitor indicating that EMERGENCY STOP
has been actuated.
â—† Unlock the EMERGENCY STOP button again.
â—† Press one of the keys for moving the lifting column.
– The lifting movement is enabled again.
Maintenance plan for checking the system 0
The tests and inspections required by national laws or regulations, such as
DHHS regulations or RöV (constancy tests), are not part of the activities listed in
this maintenance plan.
If national laws or regulations specify more frequent checking and/or mainte-
nance, this must be observed.
Maintenance work should be performed by trained technical personnel only.
To keep the system in an optimum condition, we recommend that you conclude
a maintenance contract. In the case of questions relating to maintenance/mainte-
nance contract please contact our Siemens Customer Service.
Please observe the relevant information in
(→ Register 2: Safety, Page 5).
Checking the system
The stated functions are minimum requirements.
Procedures to be performed Function Interval
Mechanical safety Damage to the housing, system movements and
options (e.g. spacer)
12 months
Electrical safety Ground conductor, grounding strap on the C-arm sys-
tem, cables and plugs
12 months
Functional check Emergency stop and lifting column 12 months
Image quality Image display and image processing 12 months