Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02 17
of 34
Current/voltage values
Generation of radiation 0
Generator 0
Voltage Long-term current consumption Short-term current consumption
100 V
110 V
120 V
127 V
200 V
230 V
240 V
17 A
16 A
15 A
14 A
9 A
8 A
8 A
29 A
26 A
24 A
23 A
15 A
13 A
12 A
Pulsed Fluoroscopy Min. pulse width 7 ms
Pulse rate 15 f/s/23 mA
Max. pulsed output 2.3 kW
Fluoroscopy 40 kV to 110 kV/0.2 to 15.2 mA (max. 1000 W)
Digital Radiography 40 kV to 110 kV/0.2 to 23 mA (max. 1000 W)
Current-time-reference
product
1.6 mAs/102 kV
Tolerances kV ± 10% (measured with spectrometric kV method)
mA ± 8% ± 0.1 mA (measured in rectified high-voltage circuit)
Fluoroscopic time 1 digit (6 s) ± 5%
Power rating 1.4 kW (102 kV/1.6 mAs)
Min. pulse width 7 ms
Max. pulsed output 2.3 kW
Exposure times min. 50 ms at 40 kV 20 mA
max. 10 s

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors