EasyManua.ls Logo

Siemens ARCADIS Varic - Cassette Holder; Dose Measuring Chamber (DAP Meter); Options

Siemens ARCADIS Varic
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02 21
of 34
Options 0
Dose measuring chamber (DAP meter) 0
Specifications
Work environment
Cassette holder 0
Technology Ionization chamber
Active area Diameter 90 mm, area 63.2 cm
2
DAP resolution 0.01 µGym
2
/s equivalent to cGycm
2
/s
Maximum
measurable DAP
3000.00 µGym
2
/s equivalent to cGycm
2
/s
Measuring inaccuracy < 1% under constant pressure and temperature
Energy range 40 - 125 kV ± 6%
Attenuation equivalent
(Inherent filtration)
< 0.5 mm Al equivalent
Dose rate linearity better than ± 2%
Measuring readiness 10 seconds after power on
Temperature +10 °C to +70 °C
Relative humidity 15% to 75% (without condensation)
Barometric pressure 700 hPa to 1060 hPa
Format 24 cm x 30 cm (10 inches x 12 inches)
Format with grid Pb r17 N70, f
o
85

Table of Contents

Other manuals for Siemens ARCADIS Varic

Related product manuals