Operator Manual
10
of 28 SPR2-310.620.30.02.02
Safety
Electrical faults 0
In the case of danger for patients and operators (e.g. if there is no live image on
the monitor but the radiation indicator is lit) or danger for the product, you must
disconnect the power plug immediately. The ARCADIS Varic will be shut down
completely and disconnected from the power supply. This will
❏ switch off radiation
❏ abort the current system program
❏ abort and cancel current operating sequences
❏ delete all image information not saved to a hard disk
Only after the cause of the hazard has been clearly identified and remedied may
the system be reconnected to the power supply. In all other cases, e.g. system
malfunction, contact Siemens Customer Service immediately.
System failure 0
The user must have a replacement unit available if a system failure could predict-
ably cause a critical situation resulting in patient injury during a medical examina-
tion.
Caution
Electrical faults
Risk of injury to the patient and operating personnel.
◆ Immediately disconnect the power plug from the socket.
Caution
❏ System component failure during the cardiovascular examination
❏ The footswitch has been removed and radiation release is no longer possible
Risk of injury to the patient!
◆ Ensure a replacement unit is available during the examination procedure.