Maintenance
ARCADIS Varic
SPR2-310.620.30.02.02 7
of 16
Updating virus protection 0
Virus protection should be updated every 2-3 days to make sure the
ARCADIS Varic has the latest virus definitions installed. The update process
takes a few minutes and should be performed during a break or outside of exam-
ination hours.
Calling up the dialog
window
To update the virus protection, open the SoftwareDistribution dialog window in
the Local Service window.
â—† Select Options > Customer Service > Local Service in the main menu of the
task card.
–The Service Software dialog box is displayed.
â—† Delete the default code in the left window of the entry area Please enter
password and click OK.
– The start window for service settings is displayed.
â—† Click SoftwareDistribution.
–The SoftwareDistribution window appears.