Technical Data
ARCADIS Varic
SPR2-310.620.30.02.02 9
of 34
Dosimetric information 0
3rd edition IEC 60601-2-54 Air kerma strength 0
The skin dose values were measured at a distance of 30 cm from the image
intensifier input with a 20 cm PMMA phantom pursuant to IEC 60601-2-54:2009,
203.5.2.4.5.101.
*) standard examination program provided
Program* Curve Format Dose Air kerma reference
CFC HC2 Normal medium 19 mGy/min
Standard medium
push
20 mGy/min
CFC IOD Normal low 12 mGy/min
Standard medium 18 mGy/min
high 22 mGy/min
Card IOD Normal high push 33 mGy/min
Program* Curve Format Noise
reduction
Air kerma reference
DR DR
1000
Normal k = 4 0.2 mGy/f (image)
k = 8 0.4 mGy/f (image)
k = 16 0.8 mGy/f (image)
k = 32 1.6 mGy/f (image)