Operator Manual
10
of 34 SPR2-310.620.30.02.02
Technical Data
Maximum air kerma strength 0
The skin dose values were measured at maximum generator power at a distance
of 30 cm from the image intensifier input pursuant to IEC 60601-2-54:2009,
203.5.2.4.5.101.
*) standard examination program provided
**) In ADR stop mode, the maximum kV value is set
Program* Curve ADR stop** Air kerma reference
CFC HC2 110 kV/5 mA 52 mGy/min
Standard HC1 110 kV/3 mA 32 mGy/min
Program* kV mAs Air kerma reference
Cassette 110 69 12 mGy/f (image)
Exposure 72 mGy/min