Operator Manual
12
of 34 SPR2-310.620.30.02.02
Technical Data
Guidelines and manufacturer's declaration –
Electromagnetic emissions 0
The system is intended for use in an electromagnetic environment as specified
below. The customer or the user of the system should ensure that it is operated
in such an environment.
The device or system should not be operated standing next to or on/under other
devices. If adjacent or stacked use is necessary, the equipment or system should
be observed to verify normal operation in the configuration in which it will be
used.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions
according to CISPR 11
Group 1 The system uses RF energy exclusively for its internal function.
Therefore, its RF emissions are very low and not likely to cause
any interference in nearby electronic equipment.
RF emissions
according to CISPR 11
Class A The system is not suitable for use in domestic establishments
and establishments directly connected to a public power supply
network which supplies buildings used for domestic purposes.
Harmonic emissions
according to IEC 61000-3-2
Not applicable The system is a professionally used unit with a total rated power
larger than 1 kW. There are no limit values for this rated power.
Voltage fluctuations/
flicker emissions
according to IEC 61000-3-3
Complies with the speci-
fied regulation