Storing on multi-session media 0
As your system is configured for multi-session, you can store your data to new,
unrecorded CD/DVD-Rs, or CD/DVD-Rs that have already been written to.
The other data on the CD/DVD-R is not lost, the new data is simply added.
DICOM Viewer In the first session, a DICOM viewer is written to the CD/DVD together with the
image data. This allows you to view the images stored on the medium on any
computer. The DICOM viewer is launched directly from the medium (automati-
cally after the CD/DVD is inserted, or manually by double-clicking
\SYNGO_FV\SYNGO_FV.EXE). No installation of files takes place on the comput-
er in question.
The DICOM viewer also allows you to store imported image information in the
AVI, Bitmap and JPEG file formats. It is currently not possible to display com-
pressed data on the DICOM viewer.
Selecting data Usually you export the data to CD/DVD-R from the Patient Browser, where you
can simply select the required data objects.
Make sure that the correct CD/DVD-R is in the drive.
Open the Patient Browser.
Select the patient, study, series or images that you want to store on
CD/DVD-R in the navigation or content area.
Caution
Write error during the recording of additional sessions on the medium in multi-
session mode.
Previously stored data and data of the current session can no longer be
read!
Do not delete the data stored on a medium from the local database until you
have successfully finalized the medium and verified the legibility of the data.
Please note that CD/DVD-Rs are not suitable for long-term archiving.