Operator Manual
20
of 24 SPR2-310.620.30.02.02
Configuration
The Roadmap tab card 0
This tab card is available only with the Subtraction option.
Parameter settings You can change the following parameters on the tab card:
You have to select the setting "black" for the Bone Display parameter. Other-
wise, the system will apply incorrect image brightness values and automatic
vessel contour detection fails.
Parameters Remarks
Opacification Possible values: min, max
Dose Level Possible values: medium, high
Landmark With the slider: 0 to 30%, in steps of 5%
Bone Display Possible values: white, black
Autostore off (no automatic storage)
all (all images are stored)
Storage Rate f/s Only if Autostore = all
Setting in frames/s
Autoloop Check box for activating automatic play-
back of a sequence of images at the end
of the exposure
Parameter Set Default parameter set for automatic ves-
sel contour detection