kV/mA curves for Pulsed Fluoroscopy (> 2 frames/s) 0
(1) LD 330 W
SIREMATIC LD Low Dose
standard mode max. 21.8 mA
in Power Mode 600 W max. 22.7 mA
(2) S1 330 W
SIREMATIC S Standard
standard mode max. 22.4 mA
in Power Mode 600 W max. 23.0 mA
(3) S2 550 W
SIREMATIC S Standard
standard mode max. 23.0 mA
in Power Mode 1000 W max. 23.0 mA
(4) HC1 330 W
SIREMATIC HC High Contrast
standard mode max. 23.0 mA
in Power Mode 600 W max. 23.0 mA
(5) HC2 550 W
SIREMATIC HC High Contrast
standard mode max. 23.0 mA
in Power Mode 1000 W max. 23.0 mA
(6) IOD 550 W
SIREMATIC IOD Iodine-contrast optimized
standard mode max. 23.0 mA
in Power Mode 1000 W max. 23.0 mA