Image Processing
ARCADIS Varic
SPR2-310.620.30.02.02 43
of 78
Inverting images 0
When an image is inverted, bright areas in grayscale images are displayed dark
and dark areas are displayed bright. The gray scale assignment in the LUT
(look-up table) is inverted.
â—† Select the images that you want to edit.
â—† Select Image > Invert Gray Scale.
—or—
â—† Click the Invert Gray Scale button on the Image subtask card.
To reset the original view, select Image > Invert Gray Scale again.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors