Configuration
ARCADIS Varic
SPR2-310.620.30.02.02 15
of 24
The Fluoroscopy tab card 0
The operating programs for continuous fluoro and pulsed fluoro are configured on
the same tab card. There are some parameters that are specific to each of these
operating programs (CONTINUOUS and PULSED areas), and some that are
identical for both (COMMON area).
Parameter settings You can change the following parameter on the tab card of the individual area:
Parameters Remarks
Noise Reduction Parameter sets for noise reduction and
or motion detection are possible
Noise Reduction Low Noise Reduction Low is always less or
equal to noise reduction
Noise Reduction LIH Special noise reduction factor for
LIH image to guarantee appropriate
image quality even with short-time
release
off (no noise reduction)
k = 2, k = 4, k = 8 (noise reduction on
with selected k factor); with PFC for
pulse rates > 2 p/s only

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors