Image Processing
ARCADIS Varic
SPR2-310.620.30.02.02 75
of 78
Exporting Sub/Roadmap image data 0
In the Exporting card you can define the data of Subtraction/Roadmap examina-
tions that are to be saved during exporting and transfer (Transfer menu).
Click the Exporting card into the foreground.
Click one or more of the following options:
export for Subtraction Image Display
Subtracted images are saved. These can also be viewed in non-Siemens
systems. Subtraction functions can no longer be performed with these
data.
export for Native Image Display
Native images are saved. These can also be viewed in non-Siemens sys-
tems.
export as ’Raw Data’
The original data set is saved. It can be post-processed in other ARCADIS
systems or in systems with subtraction functions.
Native Image Display is the default setting for export.
Check the corresponding check box, if applicable.
A dialog box is shown for every transfer of Subtraction/Roadmap images.
This allows you to select data that differ from the default settings.