EasyManuals Logo

Siemens ARCADIS Varic Operator's Manual

Siemens ARCADIS Varic
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #296 background imageLoading...
Page #296 background image
Operator Manual
16
of 98 SPR2-310.620.30.02.02
Documentation
Automatic
exposure on
This function may accelerate the workflow associated with evaluation. However,
it is recommended to switch it off prior to the next examination.
â—† Select Options > Auto Expose on the right monitor.
Or
â—† Click the Auto Expose button.
– A film sheet that is completely filled is automatically printed or exposed as
soon as an image for the next (empty) film sheet is transferred.
If the last film sheet is only partially filled, then the configuration settings deter-
mine when it is sent to be filmed.
(→ Page 54)
Automatic
exposure off
â—† Deselect Options > Auto Expose.
Or
â—† Click the active Auto Expose button.
Aborting automatic exposure
If errors occur during automatic filming, or you notice that defective images have
been filmed/printed, you can abort automatic exposure at any time.
â—† Deselect Options > Auto Expose on the right monitor.
â—† Eliminate the fault.
— or —
â—† Delete the faulty images from the virtual film sheet.
â—† Activate Auto Expose again.
— or —
â—† Transfer the remaining sheets to the camera/printer manually.

Table of Contents

Other manuals for Siemens ARCADIS Varic

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens ARCADIS Varic and is the answer not in the manual?

Siemens ARCADIS Varic Specifications

General IconGeneral
TypeMobile C-arm
Image Intensifier9-inch or 12-inch
DICOM CompatibilityYes
Frame RateUp to 30 fps
Focal Spot Size0.3/0.6 mm
MonitorDual 18-inch TFT LCD monitors