TECHNICAL INFORMATION, Patient Data Text Format
Sysmex SF-3000 Operator's Manual -- Revised September 1995 C-11
(n) IP Message Information indicates if the WBC, RBC, and PLT parameters are
flagged with Abnormal or Suspect IP Messages.
"0": No flag.
"1": Flagged.
PLT Suspect
PLT Abnormal
RBC Suspect
RBC Abnormal
WBC Suspect
WBC Abnormal
(o) "X" in the table indicates the numeric value and "F" in the table indicates the flag
code. Zero suppression is not performed. The flag code indicates the following:
"0": The analysis data is within the preset Mark Limits.
"1": The analysis data is greater than the preset Upper Mark Limit.
"2": The analysis data is less than the preset Lower Mark Limit.
"3": The analysis data exceeds the linearity limit.
"4": The analysis data is less reliable due to the abnormality of the blood sample
itself.
Most Significant Digit
Least Significant Digit
Flag
Data
(p) In case that the analysis data is reported "****", "++++" or " - - - -" with overflow
error, with analysis error, or with some flagging, an asterisk mark (*) will be
placed in front of the most significant digit and remaining four or five digits will be
turned to 0 (zero), thus "*0000" will be sent for the analysis result. If the sample
was not analyzed according to the order information, four or five blank spaces will
be sent " ".
(q) Reserve 1 through 12 -- (Not used)