INSTALLATION, CP Sampler Unit
Sysmex SF-3000 Operator's Manual -- Revised September 1995 B-15
6.3 Connection to Main Unit
(1) Using a screwdriver, remove the dummy cover from the fluid connector.
(2) Insert the sampler about halfway onto the sampler mounting rail.
(3) Connect the tubes and cables to
the sampler connection unit at the
bottom of the Main Unit (see
Figure B-19).
Fluid connector 1
Silicone tube 1
Cord and cables 4
(Including ID Unit Cable)
When attaching the fluid
connector, make sure to attach the
rubber seal and align protrusions.
Grounding
Wire
ID Unit Cable
Figure B-19: Connect Tubes and Cables
CAUTION: When assembling the sampler, connect the ground cable
(green/yellow cable) first.
When disassembling the sampler, disconnect the ground cable
last.
(4) Reinstall the Main Unit front panel.
(5) Push the sampler unit all the way toward the Main Unit. Make sure that the
Sampler Unit is fixed securely with the catchers.
Front Panel
Figure B-20: Install the Sampler