7.2 Execute RBC Aperture Clog Removal Sequence
If the error message "RBC Count Time Too Long" or "RBC Slow To Start" appears on
the LCD, execute a clog removal sequence as described below.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
Sampler
XM HC GP DP SU
Maintenance
Reagent
Clog
Removal
Drain
Sample
Service
Sequence
Test
Service
Data
More
Return
Figure 5-7-5: Maintenance Second Menu
(2) Press the [Clog Removal] key to execute the clog removal sequence.