EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #423 background imageLoading...
Page #423 background image
ADJUSTMENT, Main Unit
9-6 Sysmex SF-3000 Operator's Manual -- Revised September 1995
4.1 Adjusting to 0.6 kg/cm
2
This pressure is supplied to the sheath detector for WBC analysis.
CAUTION: Before adjusting the 0.6 kg/cm
2
, check the 2.0 kg/cm
2
pressure and
adjust if necessary.
(1) Display the 0.6 kg/cm
2
pressure on the LCD (refer to Section 2.2 in this chapter).
(2) Pull out the adjustment knob for the 0.6 kg/cm
2
regulator to release the lock.
Increase
Adjustment
Knob
Decrease
Pull out
and turn it.
Figure 9-7: 0.6 kg/cm
2
Regulator
(3) Turn the knob until the pressure is within 0.6
± 0.005 kg/cm
2
. Turning the knob
clockwise increases the pressure.
CAUTION: Always adjust the pressure from a level that is below the target. If the
pressure is high, decrease it all at once to a value that is below the
target; then gradually increase the level to the target.
NOTE: The required range for normal operation is within 0.6 ± 0.03 kg/cm
2
.
(4) Press the knob in. This locks the regulator.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals