Loading...
RESULT INTERPRETATION/REPORTS, Stored Data, List Display
Sysmex SF-3000 Operator's Manual -- Revised September 1995 4-13
3.2 CBC8 Screen (Page 2)
This screen lists the analysis data for 8 CBC parameters: WBC, RBC, HGB, HCT,
MCV, MCH, MCHC and PLT.
To switch to this screen from "P1. Sample Info.", press [] key. To switch to this
screen from "P3. WBC+5Diff#", press [] key.
900/1000(1000)
Search
Select
Display
OutputSort Valid.
More Return
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Next
Prev
Mark
Stored Data =ALL SAMPLES =
SAMPLE NO. WBC RBC HGB HCT MCV MCHC PLTMCH
123-56789-101*
123-56789-101*
123-56789-102
123-56789-103
123-56789-104
123-56789-105
123-56789-106
123-56789-107
123-56789-108
123-56789-109
123-56789-110
123-56789-111
123-56789-112
123-56789-113
123-56789-114
123-56789-115
123-56789-116
123-56789-117
123-56789-118
123-56789-119
5.38
7.68
4.82
5.90
1.41
4.68
1.06
11.79
5.20
6.60
6.21
6.73
8.44
4.37
23.31
6.38
5.26
6.26
4.94
6.04
3.90
3.65
3.99
4.12
2.78
4.47
2.59
3.91
3.66
2.94
4.29
420
1.74
3.94
5.22
2.82
3.64
4.72
3.51
4.68
11.6
12.1
12.0
11.6
8.8
14.4
10.2
12.9
11.4
8.6
13.9
12.4
7.0
13.5
16.1
9.0
11.6
14.4
12.2
14.5
34.8
35.3
36.1
36.5
25.6
44.3
28.9
36.7
34.1
27.2
41.5
36.1
19.9
39.8
49.6
27.0
34.6
45.0
34.0
44.4
89.2
96.5
90.5
88.6
91.9
99.2
111.7
93.9
93.4
92.5
96.8
85.9
114.2
101.0
95.0
95.9
95.1
95.4
96.9
94.9
29.8
33.2
30.1
28.2
31.8
32.4
39.7
33.1
31.4
29.4
32.5
29.8
40.4
34.5
30.9
32.0
32.0
30.7
35.1
31.1
33.4
34.4
33.2
31.8
34.6
32.7
35.5
35.3
33.6
31.8
33.6
34.7
35.4
34.2
32.6
33.4
33.6
32.2
36.2
32.7
323
223
271
172
163
348
31
242
445
281
498
297
54
190
389
308
476
257
79
263
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
v
Graphic
Figure 4-3: CBC8 Screen (P2)
1) "v" (validated) Mark
Indicates that this data is validated. Refer to Section 5: Data Validation in this
chapter.
2) SAMPLE NO.
Sample No., as read by bar code reader or input by the operator.
3) Indications
The following indications appear to the right of analysis data:
! (backlit in red):
Indicates that the data is outside the Critical Limits.
* (backlit in red):
Indicates that the data is outside the Linearity Limits.
+, – (backlit in red):
Indicates that the data is outside the Sampler Stop Limits.
* (backlit in yellow):
Indicates that the data is unreliable because of an abnormality detected in the
sample.
+, – (backlit in yellow):
Indicates that the data is outside the Mark Limits.
NOTE: The above indications are listed on the priority basis.