MAINTENANCE & SUPPLIES, Monthly
5-16 Sysmex SF-3000 Operator's Manual -- Revised July 1996
5.2 Execute Clean Waste Chamber Sequence
Clean the waste chambers and trap chambers once a month, or every 2500 samples, and
reset the counter as described below.
NOTE: After every 2500 cycles or a month after the previous cleaning, the
message "Clean the Waste Chamber" will be displayed at power ON.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
Sampler
XM HC GP DP SU
Maintenance
Reagent
Clog
Removal
Drain
Sample
Service
Sequence
Test
Service
Data
More
Return
Figure 5-5-4: Maintenance Second Menu