CALIBRATION
Sysmex SF-3000 Operator's Manual -- Revised September 1995 7-3
2. HGB/HCT AUTO
Using five to ten samples of fresh normal whole blood, perform calibration of the
SF-3000 including the following four steps:
1. Enter reference values. 3. Review and Select data for editing.
2. Analyze samples. 4. Update calibration values.
2.1 Overview of the HGB/HCT Auto Program
(1) Press [More] to change the displayed root menus, then [Calib.] keypad. The
Calibration screen containing submenu will appear as shown in Figure 7-1.
Cal.
History
Return
DP:1234567890123
Sysmex
Ready
HGB/HCT
Auto
HGB/HCT
Manual
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Calibration
Figure 7-1: Calibration Screen
(2) Press [HGB/HCT Auto]. The message "Input Password" and numeric keypads
will appear. Input the password and press [ENTER].
CALIB.
DP:1234567890123
Sysmex
Ready
ENTER
C
78
9
456
123
0
Input Password
Password
***
CANCEL
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Return
Cal.
History
HGB/HCT
Manual
HGB/HCT
Auto
Figure 7-2: Password Input Screen