INSTALLATION
Sysmex SF-3000 Operator's Manual -- Revised September 1995 B-27
8.5 Waste Container
Connect one end of the polyurethane tube provided (6 mm ID x 9 mm OD) to the
instrument waste outlet nipple (marked WASTE). Connect the other end to the waste
sewer or waste container.
Waste Container
Waste Outlet Nipple
Figure B-38: Waste Tube Connection
CAUTION: Position the waste sewer or waste container lower than the instrument
waste nipple. The length of the waste tube should not exceed ten
meters (10 m).