CALIBRATION
7-6 Sysmex SF-3000 Operator's Manual --- Revised September 1995
2.2 Setting (Entering Reference Values)
To perform HGB/HCT Auto calibration program, you must enter, as the reference
values, the measured HGB and HCT values obtained using reference methods. A
maximum of 10 sets of data can be registered.
(1) From the HGB/HCT Auto submenu, press [Setting]. "Reference" display will be
backlit in blue, and numeric keypads will appear.
(2) Select the "Reference" sample number by using the [↑] and [↓] keys to move the
cursor.
Select either "HGB" or "HCT" by using the [←] and [→] keys.
HGB
1
2
3
4
5
6
7
8
9
10
Reference Analyze Compensation [%]
Avg.
CALIB.
DP:1234567890123
Sysmex
Not Ready
HGB HGBHCT HCT HCT
< HGB / HCT Auto>
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Sample No. 1 HGB
0.0
89
7
4
1
0
C
5
2
6
3
ENTER
CANCEL
Return
Setting Analyze Exclude Exe. Cal. Graphic
14.5
44.3 14.6
↑
→
←
↓
Figure 7-5: HGB/HCT Auto - Setting Screen (1)
(3) Use the numerical keys to enter the HGB/HCT reference values obtained using the
reference methods. The input format should correspond with the unit settings.
Press [Enter] to move the cursor to the next item.
NOTE: Press the [C] key to delete the entered values.