MAINTENANCE & SUPPLIES, As Needed
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-41
7.5 Execute WBC Flow Cell Air Bubbles Removal Sequence
When the scattergrams show abnormally enlarged groups of cells, air bubbles may exist
in the WBC Flow Cell. At this time, remove the air bubbles as described below.
NOTE: No WBC IP message or no analysis error may occur if air bubbles
exist in the WBC Flow Cell.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
Sampler
XM HC GP DP SU
Maintenance
Reagent
Clog
Removal
Drain
Sample
Service
Sequence
Test
Service
Data
More
Return
Figure 5-7-19: Maintenance Second Menu