SYSTEM SETUP
Sysmex SF-3000 Operator's Manual -- Revised September 1995 11-27
4.1 WBC
Use this program to set IP message limits for WBC.
To set abnormal limits for WBC:
(1) Press [WBC] key in the Flag System submenu.
The WBC setting screen containing the current settings for the various IP messages
and numeric keys will appear.
(2) Use the [↑] and [↓] keys to select the desired abnormal message and setting item.
(3) Use the numeric keys to input the new limit value. Press [ENTER] to input the
values and move to the next item.
WBC RBC PLT Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
ABNORMAL
COMMENT
NEUT#
Neutropenia
3
ITEM
ENTER
QUIT
C
78 9
456
123
0
1.00
WBC
NEUT# [X10 /uL]1.00
NEUT%
Neutrophilia
[%]
Lymphopenia
Lymphocytosis
Monocytosis
Eosinophilia
Basophilia
Leukocytopenia
Leukocytosis
0.0
11.0
100.0
0.80
0.0
4.00
100.0
1.00
100.0
0.70
100.0
0.20
100.0
2.50
18.00
3
[X10 /uL]
[%]
3
[X10 /uL]
[%]
3
[X10 /uL]
[%]
3
[X10 /uL]
[%]
3
[X10 /uL]
[%]
3
[X10 /uL]
[%]
3
[X10 /uL]
3
[X10 /uL]
NEUT#
NEUT%
LYMPH#
LYMPH%
LYMPH#
LYMPH%
MONO#
MONO%
EO#
EO%
BASO#
BASO%
WBC#
WBC#
SET
VALUE
•
↑
↓
Figure 11-21: WBC Flag System Setting Window
NOTE: The factory settings are given in Table 11-2.