QUALITY CONTROL
Sysmex SF-3000 Operator's Manual -- Revised December 1995 6-41
8. MANUAL QC ANALYSIS
X Control and L-J Control can be performed in the Manual mode, Closed mode, or
Manual CP mode using control blood material. The overview of the QC analysis in these
three modes is described in this section. Refer to Chapter 2, Section 3: QUALITY
CONTROL ANALYSIS for step-by-step procedures to run QC in each mode.
(1) Make sure that the instrument is in the Ready status.
(2) Select the analysis mode, Manual or Closed.
(3) Specify the file No. in which the QC data are to be stored.
(4) Verify that the control lot information is correctly set for the selected QC file, if
necessary.
(5) Analyze the control material in manual (open or closed) mode. Refer to Chapter 2,
Section 4 for the procedures.
(6) After the analysis is completed, the results are displayed and stored in the memory.
NOTE: The analysis results up to 1000 QC samples are stored in the
memory.
• When L-J Control program is used:
The QC analysis results are displayed as shown in Figure 6-38a. The sample No.
on the graphic display is indicated as "QC01" (see Figure 6-38c).
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No. 1234567890123
Sampler
JudgmentData
WBC
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
QC01 < L-J Control >
NEUT#
LYMPH#
MONO#
EO#
BASO#
NEUT%
LYMPH%
MONO%
EO%
BASO%
GRAN-X
GRAN-Y
WBC/BA-X
WBC/BA-Y
GRAN-Y(W)
Graphic Return
7.23
4.74
14.1
41.6
87.8
29.7
33.9
46.4
15.7
213
9.8
10.3
24.3
4.62
1.55
0.71
0.18
0.17
63.9
21.4
9.8
2.5
2.4
177.4
177.2
44.2
94.7
43.0
JudgmentData
Figure 6-38a: QC Data Display Screen (L-J Control)