INSTALLATION
Sysmex SF-3000 Operator's Manual -- Revised August 1998 B-17
7. MANUAL CP UNIT (Optional)
7.1 Main Unit
NOTE: • When the Manual CP Unit is installed, the Wiring Cord
No.1777 (P/N 943-2801-1) is necessary.
• When the Manual CP Unit is installed, the grounding wire and
some screw bindings are necessary. Refer to
Section 1.2
and
order the parts if needed.
• When the ID reader installation is not required, install the ID
reader dummy cover with the grounding wire on the back of the
front panel. Refer to
Section 3.2
in this chapter.
(1) Open the front cover and remove the front panel. See Section 3: Remove Packing
Material in this chapter.
WARNING! When working with the front cover open, make sure that the
stop bar is in place.
(2) Install the KCP-4 dummy cover
using the M3x6 screws.
(3) Attach the grounding wire to the
KCP-4 dummy cover with the M4
screw.
(M4)
Figure B-22: Attach Grounding Wire
(4) Remove four fixing screws and
open the LCD Unit. Then PCB
No.1234 will appear.
SCREWS
SCREWS
Figure B-23: Open LCD unit