SYSTEM SETUP
11-22 Sysmex SF-3000 Operator's Manual -- Revised September 1995
3.4 Sampler Stop Condition
Use this program to set the sampler (auto mode) analysis stop conditions.
The SF-3000 will stop sampler analysis when an error occurs if the conditions for the
error are enabled in the sampler stop conditions program.
To set sampler stop conditions:
(1) Press [Stop Cond.] key in the Data Criteria submenu.
The Stop Conditions setting screen containing the current setting conditions will
appear.
< AUTO MODE STOP CONDITIONS >
Aspiration ERR
Inadequate Sample
X M Limit ERR
L-J Limit ERR
ID Read ERR
Rack ID Read ERR
Low Count ERR
Stop Limit ERR
Manometer ERR
Mark
Limits
Stop
Limits
Stop
Cond.
Critical
Limits
Return
Quit
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Control Expired
√
√
√
√
√
√
√
√
√
√
Figure 11-16: Sampler Stop Conditions Setting Window
NOTE: The conditions illustrated on the screen above are the factory settings.