RESULT INTERPRETATION/REPORTS
4-2 Sysmex SF-3000 Operator's Manual -- Revised February 1997
2. INTERPRETIVE (IP) MESSAGES
The SF-3000 analyzes, displays and prints hematology information in a format designed
to aid in the separation of POSITIVE or NEGATIVE data results. All analyzed samples
without analysis errors can be separated into a POSITIVE and NEGATIVE category
according to preset criteria.
The system bases its judgment on comprehensive surveys of numerical data, particle size
distribution, scattergrams, etc., and provides easy to understand flags/messages
indicating the instruments findings. These flags/messages are referred to as "IP
(Interpretive Program) messages."
[x10 /uL]
[x10 /uL]
1.2
[x10 /uL]
XM
6.94
123-56789-101
List
Edit
Info.
DeleteOutput
Valid.
NO.
R B C P L T
D I F F W B C / B A S O
[g/dL]
[fL]
WBC
NEUT
LYMPH
MONO
EO
BASO
RBC
HGB
HCT
MCV
MCH
MCHC
RDW-SD
RDW-CV
PLT
PDW
MPV
P-LCR
250fL 40fL
RBC Flag
PLT Flag
WBC Flag
DP:1234567890123
Sysmex
Ready
Sampler
Return
3
[%]
[%]
[%]
[%]
[%]
6
[g/dL]
[%]
[%]
[%]
[fL]
[fL]
[fL]
3
[pg]
Manual Mode Next No.1234567890123
HC GP DP SU
31.6
30.8
3.8
2.4
0.8
4.51
***.*
42.9
95.4
31.3
32.9
49.3
14.7
290
11.9
10.3
27.1
45.4
44.4
5.5
3.5
POSITIVE ERROR
R:1000-10 20:48 1994/6/30
L
Thrombo
Left Shift?
Stored Data ALL SAMPLES
900/1000 (1000)
Figure 4-1: IP Message Screen
POSITIVE (red back-light) : A POSITIVE result indicates a sample is judged
abnormal according to preset criteria for analysis
numerical values and cell morphology.
NEGATIVE (green back-light) : A NEGATIVE result indicates that the sample is
normal, i.e., has no analysis errors or IP messages.
The system sub-categorizes and flags POSITIVE results as "Diff.", "Morph." and/or
"Count" during WBC, RBC and PLT analysis.
These flags appear when abnormal cell population is
detected during computer analysis of the particle size
distribution, scattergrams and 23 parameters.
Diff. : Indicates abnormality in the WBC differential parameters.
Morph. : Indicates abnormal cell morphology.
Count : Indicates abnormality in the blood cell numerical count.
These sub-categories appear below the POSITIVE flag.