WORK LIST
3-2 Sysmex SF-3000 Operator's Manual --- Revised July 1996
2. EXECUTING A WORK LIST PROGRAM
Press [Work List] keypad in the Root Menu. The Work List as shown in Figure 3-1
appears. The analysis information is input using this display.
When the Work List screen appears, the second menu also appears in the bottom line.
The second menu provides the following processing modes: "Down Load", "Output" and
"Delete". These modes can be selected through the LCD touch panel. Refer to the
following sections for the procedures.
[0/15]
Work
RACK-TUBE
OrderRack Repeat
Output Delete
Down
Load
Prev
Next
Sample
No.
SAMPLE No. ORDER
List
ENTER
C
78
9
456
123
0
-
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
0002-05
Rack No.
0001-01
0001-02
0001-03
0001-04
0001-05
0001-06
0001-07
0001-08
0001-09
0001-10
0002-01
0002-02
0002-03
0002-04
123-456-789-10
123-456-789-11
123-456-789-12
123-456-789-13
123-456-789-14
123-456-789-15
123-456-789-16
123-456-789-17
123-456-789-18
123-456-789-20
123-456-789-21
123-456-789-22
123-456-789-23
123-456-789-24
↑
↓
CANCEL
0002-05 123-456-789-25
Mark
Figure 3-1: Work List Display
2.1 Work List Display
The Work List display contains the following information:
Displayed Items
• RACK-TUBE
Displays the rack number (1 - 9999) and the tube position number indicates the
location of the blood sample tube in the rack (1 - 10).
• SAMPLE No.
Displays the sample ID number (up to 13 digits).
• ORDER
Displays the analysis order inquiry and the status of the sample analysis in progress.
Blank : Indicates that the sample is not to be analyzed.
Circle ( ) : Indicates that the sample is to be analyzed but has not
been analyzed.
Double Circle ( ) : Indicates that analysis is complete.
Cross (X) : Indicates that an analysis error occurred, preventing
the completion of sample analysis.
• [#/##]
Number of marked samples/Number of registered samples in the list is indicated
above [Repeat] keypad.