EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #180 background imageLoading...
Page #180 background image
MAINTENANCE & SUPPLIES, Weekly
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-9
4.2 Execute Clean WBC Detector Sequence
Clean the WBC flow cell and the reaction chambers once a week, or every 600 samples,
and reset the counter as described below.
NOTE: After every 600 cycles or a week after the previous cleaning, the
message "Clean the WBC Detector" will be displayed at power ON.
WARNING! The WBC reaction chambers should be considered potentially
biohazardous. Do not handle the chambers without proper protective
equipment. Wear rubber gloves when executing this procedure.
Wash your hands with an antiseptic solution after completing the
procedure.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
Sampler
XM HC GP DP SU
Maintenance
Reagent
Clog
Removal
Drain
Sample
Service
Sequence
Test
Service
Data
More
Return
Figure 5-4-3: Maintenance Second Menu
Sysmex SF-3000 Operator's Manual -- Revised March 1996

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals