MAINTENANCE & SUPPLIES, Weekly, Clean WBC Detector
5-10 Sysmex SF-3000 Operator's Manual -- Revised March 1996
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-4-4: Service Sequence Submenu
(3) Press [Clean WBC Detector] key. The message "It will take approx. 5 min. Press
Start Switch." appears.
Pressing [Cancel] cancels the program without cleaning and returns to the Service
Sequence submenu display.
Sysmex
Ready
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
< Service Sequence >
<Clean WBC Detector>
It will take approx. 5 min.
Press Start Switch.
Maintenance
Cancel
Return
Figure 5-4-5: Clean WBC Detector Execute Message