MAINTENANCE & SUPPLIES, As Needed
Sysmex SF-3000 Operator's Manual -- Revised September 1995 5-43
7.6 Execute Replace Sheath Flow Reagent Sequence
When the WBC background count becomes abnormally high, the internal reservoir
chamber for WBC sheath flow reagent (CELLPACK) may contain air bubbles. At this
time, remove the air bubbles as described below.
(1) Press the [Mainte.] key from the root menu. The Maintenance second menu will be
displayed.
NOTE: If the [Mainte.] key is not displayed on the LCD, press [More] to
change the displayed root menus.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
Sampler
XM HC GP DP SU
Maintenance
Reagent
Clog
Removal
Drain
Sample
Service
Sequence
Test
Service
Data
More
Return
Figure 5-7-22: Maintenance Second Menu