MAINTENANCE & SUPPLIES, As Needed, Replace Sheath Flow Reagent
5-44 Sysmex SF-3000 Operator's Manual -- Revised September 1995
(2) Press [Service Sequence] key. The Service Sequence submenu will appear.
Sysmex
Ready
DP:1234567890123
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
Maintenance
Clean
Waste Chamber
Clean
WBC Detector
Clean
RBC Transducer
Replace Sheath
Flow Reagent
<Service Sequence>
WBC Flow Cell
Air Bubble Removal
Return
Figure 5-7-23: Service Sequence Submenu
(3) Press [Replace Sheath Flow Reagent] key. During the replacing sequence, the
message shown below will appear on the LCD.
Sysmex
DP:1234567890123
Maintenance Next No. 1234567890123
XM HC GP DP SU
Sampler
< Service Sequence >
Replacing Sheath Flow Reagent.
Please Wait.
Maintenance
Running
Clean
RBC Transducer
Return
Figure 5-7-24: Replacing Sheath Flow Reagent Message
(4) After the replacing sequence is completed, the instrument returns to the Ready
status.
(5) Keep the record on the maintenance check list.