CALIBRATION, HGB/HCT Manual
7-16 Sysmex SF-3000 Operator's Manual --- Revised September 1995
3.2 Executing the HGB/HCT Manual Program
(1) Press [Calib.] keypad in the root menu. The Calibration screen containing
submenu will appear.
NOTE: If the [Calib.] keypad is not displayed on the LCD, press [More] to
change the root menu display.
(2) Press [HGB/HCT Manual]. The message "Input Password" and numeric keypads
will appear. Input the password and press [ENTER].
(3) The current compensation factor is displayed. Use the [↑] and [↓] keys to position
the cursor on the desired compensation factor.
98.4 %
Edit
Calibration
HGB
HCT
Cancel
100.0 %
Return
Cal.
History
HGB/HCT
Manual
HGB/HCT
Auto
Set
CALIB.
< HGB/HCT Manual >
DP:1234567890123
Sysmex
Not Ready
Manual Mode Next No. 1234567890123
XM HC GP DP SU
Sampler
↑
↓
Figure 7-16: HGB/HCT Manual Calibration Screen (1)