CALIBRATION, HGB/HCT Manual
Sysmex SF-3000 Operator's Manual -- Revised September 1995 7-17
(4) Press [Edit] key and the numerical keys are displayed. Enter the calculated
compensation factor with the numerical keys.
98.4 %
X
M
HGB
HCT
DP:1234567890123
Sysmex Sampler
100.0 %
98.4
HGB Cal
Manual Mode
Next No.1234567890123
C
78 9
456
123
0
HC GP DP SU
ENTER
QUIT
Return
Cal.
History
HGB/HCT
Manual
HGB/HCT
Auto
CALIB.
Calibration
< HGB/HCT Manual >
Not Ready
•
↑
↓
Figure 7-17: HGB/HCT Manual Calibration Screen (2)
Pressing [ENTER] moves the cursor to the next compensation factor.
Pressing [QUIT] closes the numeric key window and returns to the previous screen
(shown in Figure 7-16).
(5) Press [Set] to update the compensation factor and end the program.
Press [Cancel] to end the program without updating the compensation factor.
Press [Edit] key to re-enter the calibration value(s).
(6) Re-analyze the samples used in the calibration process. Confirm that the measured
values do not differ significantly from the reference values and that they are within
the allowable range.
If the HGB/HCT value is consistently higher or lower than the reference value,
recalibrate. If the difference continues to exceed the limits, check for sample
abnormalities such as coagulation problems, abnormal morphology, patient
medication, old blood samples, etc. If no sample abnormalities are discovered,
contact your Sysmex service representative.