CALIBRATION
7-8 Sysmex SF-3000 Operator's Manual --- Revised September 1995
2.3 Analyze (Analyzing)
The SF-3000 uses the HGB/HCT Auto submenu program "Analyze" to analyze the
reference samples in the Manual mode.
(1) From the HGB/HCT Auto submenu, press [Analyze]. "Analyze" display will be
backlit in blue, the main unit READY LED will light and the system will be ready
to process samples.
(2) Use the [↑] and [↓] keys to position the cursor on the sample number to be
analyzed.
DP:1234567890123
Sampler
Manual Mode
HGB
1
2
3
4
5
6
7
8
9
10
Reference
Analyze
Compensation (%)
Avg.
Setting Analyze
Exclude
Graphic Return
CALIB.
Sysmex
Ready
HGB HGBHCT HCT HCT
Exe. Cal
< HGB/HCT Auto >
Next No.1234567890123
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
16.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
40.0
15.9
15.9
39.9
39.940.0
100.1
100.1 100.0
100.0
[ Current ]
HGB 98.1%
HCT 100.1%
X
M
HC GP DP SU
↑
↓
Figure 7-8: Sample Analysis Screen (1)