TROUBLESHOOTING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 8-19
Rinse Motor Function Error
Phenomena Manual rinse mechanism driving motor did not move properly, and the
movement of the manual rinse mechanism stopped.
Analysis cannot be continued until the error is resolved.
Probable Cause The resistance on the manual rinse mechanism is abnormally high.
Corrective Action 1) Check visually to make sure that no fluid connector or tubing is touching
the manual rinse mechanism upper/lower parts.
2) Check the rinse motor operation using test program. (Refer to
Section
3.3: Test Motor
in this chapter.)
3) If the error persists after the test program is completed, turn the Main Unit
power OFF, then ON again.
4) If the error still persists, contact your Sysmex service representative.
Manual Rinse Cup
Rinse Motor
Figure 8-3: Rinse Motor Location (Main Unit - Front Interior)