TROUBLESHOOTING
8-20 Sysmex SF-3000 Operator's Manual -- Revised September 1995
WB Asp Motor Function Error
Phenomena Whole blood aspiration pump driving motor did not move properly, and the
movement of the whole blood aspiration pump stopped.
An analysis error may occur, and the "Check Stored Data" message also
appears. Analysis cannot be continued until the error is resolved.
Probable Cause The resistance on the whole blood aspiration pump is abnormally high.
Corrective Action 1) Check the WB Asp motor operation using test program. (Refer to
Section
3.3: Test Motor
in this chapter.)
2) If the error persists after the test program is completed, turn the Main Unit
power OFF, then ON again.
3) If the error still persists, contact your Sysmex service representative.
Mixing Motor Speed Error
Phenomena WBC reaction chamber mixing motor does not move properly.
An analysis error may occur, and the "Check Stored Data" message also
appears. Analysis cannot be continued until the error is resolved.
Probable Cause 1) The bearing of the mixing motor is dirty.
2) The belt of the mixing motor is damaged.
3) Mixing motor is incorrectly adjusted.
Corrective Action 1) Check the motor rotation speed using test program. (Refer to
Section 3.3:
Test Motor
in this chapter.)
2) Replace the mixing motor belt. (Refer to
Chapter 5, Section 8.8: Replace
the Belt No. 6
.)
3) If the error still persists, contact your Sysmex service representative.