SYSTEM SETUP
11-24 Sysmex SF-3000 Operator's Manual -- Revised March 1996
4. FLAG SYSTEM
Use this program to set the Abnormal IP Message limits for WBC, RBC and PLT.
When analysis data of a sample is outside of the limit, the data will be flagged with an IP
message to indicate abnormal status of the data.
NOTE: The limits for the Suspect IP messages cannot be changed.
To set abnormal limits:
(1) Press [Settings] key in the root menu. The Settings second menu will be
displayed.
NOTE: If the [Settings] keypad is not displayed on the LCD, press [More]
key to change the displayed root menu.
DP:1234567890123
Auto
Mgmt.
Data
Criteria
Config.
Print
Set Value
General
Set Up
Flag
System
Return
Sysmex
XM HC GP DP SU
Ready Manual Mode Next No.1234567890123
Sampler
• Setting •
Figure 11-17: Settings Second Menu Screen