EasyManuals Logo

SYSMEX SF-3000 User Manual

SYSMEX SF-3000
605 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
SAMPLE PROCESSING
Sysmex SF-3000 Operator's Manual -- Revised September 1995 2-7
3. QUALITY CONTROL ANALYSIS
Control samples (SF CHECK) are analyzed and the data is stored in the X or L-J
quality control file. Follow the manufacturer's instructions for handling control material.
Quality control analysis data is stored in specifically designated QC files. For details see
Chapter 6: Quality Control.
NOTE: Using the X Control method, the control sample is analyzed twice
in succession and the average data point is used. The L-J Control
method, on the other hand, uses the result from one analysis as one
control data point.
3.1 QC Analysis: Manual Mode
Follow the procedure below to perform QC analysis in Manual mode. The instrument
will be in Manual mode after the power is turned ON.
(1) Make sure that the READY LED is lit in green.
(2) Press the Next No. keypad in the top line of the LCD. The Sample No. Setting
screen shown below will appear.
Sysmex
DP:1234567890123
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
Manual Sample Number
CANCEL
ENTER
C
3
9
6
0
1
7
4
2
8
5
Ð
QC
1234567890123
Analysis Mode
Manual Mode
Capillary Mode
Closed Mode
Figure 2-2: Sample No. Setting Screen
(3) The selected analysis mode is backlit in light-blue. Press [Manual Mode] to select
Manual mode, if needed.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SYSMEX SF-3000 and is the answer not in the manual?

SYSMEX SF-3000 Specifications

General IconGeneral
BrandSYSMEX
ModelSF-3000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals