QUALITY CONTROL, Target/Limit
Sysmex SF-3000 Operator's Manual -- Revised September 1995 6-21
6.4 Auto Setting
This program automatically sets the average of control data previously specified on the
QC Chart as the target value.
This program also sets either 2SD or 3SD as the Limit# value, or sets either 2CV or 3CV
as the Limit% value.
NOTE: The target/limit auto setting program uses the reference data (mean,
SD, CV) computed using the QC chart in "Calculate" program.
Obtain the reference data referring to
Section 7.3: Mean/SD
in this
chapter before accessing the auto setting program.
(1) Press [QC] to display QC second menu.
(2) Press [Target/Limit] keypad from the second menu.
The first page of the QC File Information List will be displayed.
Q C
FILE
Help Setting
Read
FD
INFO. RBC HGB HCT
Return
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready Manual Mode Next No.1234567890123
Sampler
XM
No. 1
No. 2
No. 3
No. 4
No. 5
No. 6
No. 7
30070107
1996/05/10
< Target / Limit List >
MCV MCH MCHC
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
TARGET
LIMIT#
465
14
465
14
13.9
0.5
13.9
0.5
39.2
1.6
84.3
3.4
29.9
1.5
35.5
1.8
39.2
1.6
84.3
3.4
29.9
1.5
35.5
1.8
Lot
Into.
↑
↓
→
←
GP
List
Figure 6-16: QC File Information List (P1: RBC)
(3) Use the [↑] and [↓] keys to move the cursor. Select the specific file in which target
and limit values are to be registered.
(4) Press [Settings] keypad. The password entry screen will appear.