SYSTEM SETUP
11-8 Sysmex SF-3000 Operator's Manual -- Revised September 1995
2.2 Auto Output Mode (Data Specification)
Use this program to specify the type of samples for auto output to the DP, GP and/or
HC.
To specify auto output data:
(1) Press [Output Mode] key in the Auto Management submenu.
(2) Enter the password and press [ENTER] key.
The Output Mode setting screen, containing the current output settings, will appear.
The conditions selected will be marked by a check mark (
).
D P G P H C
NEGATIVE
DIFF.POS.
MORPH.POS.
COUNT POS.
ERROR
QC DATA
< AUTO OUTPUT MODE >
DP:1234567890123
Sysmex
XM HC GP DP SU
Ready
Manual Mode Next No.1234567890123
Sampler
√√√
√√√
√√√
√√√
√√√
√√√
Quit
Output
Condition
Output
Mode
Auto
Erase
Auto
Valid.
Return
Figure 11-6: Output Mode Setting Window
(3) Press the appropriate key to select the output data. To disable the setting for the
options, press the keys marked "".
The data output selections are as follows:
NEGATIVE : outputs sample data without analysis errors or IP messages.
DIFF. POS. : outputs sample data with differential IP message.
MORPH. POS. : outputs sample data with cell morphology IP message.
COUNT POS. : outputs sample data with cell count IP message.
ERROR : outputs sample data with analysis error.
QC DATA : outputs data of QC Sample.
NOTE: The factory setting is "" for all items.
(4) Press [Quit] key to exit the program. The confirmation message will appear. Press
the appropriate key [Continue], [Set] or [Cancel].
Continue : To return to Output Mode setting screen so that setting can be
continued.
Set : To save the new settings. Program ends.
Cancel : To cancel the new settings. Program ends.