EasyManua.ls Logo

SYSMEX SF-3000 - Shutdown; Shutdown Procedure

SYSMEX SF-3000
605 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SAMPLE PROCESSING
Sysmex SF-3000 Operator's Manual -- Revised October 1995 2-49
5. SHUTDOWN
When instrument shutdown is performed, the detector chambers and manometers are
cleaned. You should run the instrument through a shutdown cycle at the end of each
day's analyses or at least once every 24 hours if running the instrument continuously.
5.1 Shutdown Procedure
(1) Press the [Shutdown] key from the root menu.
NOTE: If the [Shutdown] key is not displayed on the LCD, press [More] to
change the displayed root menus.
(2) The Shutdown Execute window as shown below will appear. Press [Cancel] key
to cancel the program.
Shutdown
Cancel
It will take approx. 10 minutes.
Set CELLCLEAN to the pipette
and press Start Switch.
XM DP:
Sysmex
Ready
Sampler
Shutdown Next No.
HC GP D P SU
Figure 2-39: Shutdown Execute Window
WARNING! CELLCLEAN (or 5% filtered sodium hypochlorite) is a strong alkaline
solution. Avoid contact with it. If you get it on your skin or clothing,
immediately wash it off with excess water.

Table of Contents

Related product manuals